SCHEU MINISTAR S® & BIOSTAR® Splint / Retainer Materials Sports Mouth Guard Materials Bleaching Tray Material Finishing & Accessories Compressor Plaster Trimmer Training Videos Acrylics, Screws, Clasps, Wires Acrylics Screws, Clasps & Bars Wires Clear Aligners. Connect the MiniSTAR S to the air source using the high-pressure hose. Slip one end of the hose onto the air valve (C) of the MiniSTAR S, and slip the other end over the air pressure line. Secure each end in place with a hose clamp. Note: Route the electrical cord and pressure. Hose away from the heating element.
Manuals and User Guides for SCHEU MINISTAR S. We have 3 SCHEU MINISTAR S manuals available for free PDF download: Operation Manual. SCHEU MINISTAR S Operation Manual (20 pages) Brand. The MiniSTAR comes with one FREE materials kit that includes (7) boxes of material. Choose one below and then add it to your shopping cart for discount. Materiaks Kit for Orthodontic Practices: Clear Biocryl Resin Kit (1) Biocryl - Assorted (2mm) (1) Biocryl - Pink (2mm) (1) Hardcast (.8mm) (1) Mouthguard - Assorted (3mm) (1) Mouthguard - Clear.
Product- Overview
Scheu Ministar Manual 2016
The compact pressure moulding unit MINISTAR S® equipped with scanning technology: scanner reads the coded material, allowing programming of all important parameters such as heating time, temperature and cooling time. In other words: precise pressure moulding results are reached in record time thanks to the extremely fast, patented heater and working pressure of 4.0 bar, adjustable via pressure regulator. New generation equipped with user-friendly membrane keypad and large display. All operating parameters are monitored and indicated on the display.
- Technical data
- Power: 230/115/100 V, 750 W
- Working pressure: 0.5 - 4.0 bar
- Dimensions (W × H × D): 500 × 210 × 400 mm
- Weight: 9 kgs
- 3 years warranty, 10 years availability of spare parts
Scheu Ministar Manual 2017
FURTHER INFORMATION
FAQ
No, you can only process material with a diameter of 125 mm in SCHEU-DENTAL pressure moulding machines. Make sure to use the correct material size to avoid any problems during pressure build-up. The rim of a too small blank gets pulled out from the clamping of balancing and locking ring. Thus, a precise and correct thermoforming result is not ensured.
Maybe there is a pellet on the edge of the model cup between material and clamping ring. Interrupt the thermoforming process, taking care the model cup rim is absolutely clean before closing the pressure chamber.
Leaking of air usually does not affect the thermoforming result, because the compressor is constantly supplying air. Nevertheless, you should check the following:
When working on the model platform,
- ensure the model is placed flush with the surface of the model platform. If it's not, trim the model flat or embed it into the pellets.
- sharp edges may have damaged or torn the material. If this is the case, remove the sharp edges on the model.
- the material may have moved when being clamped. In this case, clamp a new blank and restart the thermoforming process.
- the balancing ring may have got stuck. In this case, remove the balancing ring, clean, grease and reinsert it. Ensure it is freely movable!
When embedding the model into the pellets, sharp model edges may cause perforation mainly of thin material.
No, this is not possible. However, depending on the series of your pressure moulding device, you can set an individual heating time using the “clock” symbol on the keyboard of your device or by using the “+” and “-” buttons. Please refer to the manual of your device to find out which option applies to your device. The operation manual of the current series of our pressure moulding machines can be found on this site under 'Further Information' for download. Operation manuals of previous device series are available under SERVICE on top of this web site.
Description
Ministar S
The compact pressure moulding unit MINISTAR S® equipped with scanning technology: scanner reads the coded material, allowing programming of all important parameters such as heating time, temperature and cooling time. In other words: precise pressure moulding results are reached in record time thanks to the extremely fast, patented heater and working pressure of 4.0 bar, adjustable via pressure regulator. New generation equipped with user-friendly membrane keypad and large display. All operating parameters are monitored and indicated on the display.
Scheu Ministar Manual Pdf
- Technical data
- Power: 230/115/100 V, 750 W
- Working pressure: 0.5 – 4.0 bar
- Dimensions (W × H × D): 500 × 210 × 400 mm
- Weight: 9 kgs
- 3 years warranty, 10 years availability of spare parts